1What is a Bowie-Dick test?

The Bowie-Dick test, named after the two Scottish microbiologists who described it in 1963, is a Type 2 chemical indicator test defined in ISO 11140-1 and specified in EN 285 / ISO 17665. It does not test sterility. It tests whether a pre-vacuum sterilizer removes air from the chamber effectively and whether the steam that replaces it penetrates a porous load quickly and uniformly.

2Why air removal matters

Steam sterilizes by condensing on surfaces and transferring latent heat. Air trapped inside a pack forms a pocket that steam cannot enter, so the contents inside that pocket never reach sterilizing conditions. Pre-vacuum sterilizers pull a series of vacuums to remove air; a leaking door seal, a faulty vacuum pump or non-condensable gases in the steam supply all leave residual air behind. The Bowie-Dick test is designed to reveal exactly that.

3How to run the test

  1. Warm the sterilizer with an empty cycle at the start of the day.
  2. Place a standard Bowie-Dick test pack (or a disposable test pack that conforms to ISO 11140-4) horizontally on the lowest shelf, above the drain, in an otherwise empty chamber.
  3. Run the dedicated Bowie-Dick cycle: 134 °C for 3.5 minutes (never longer, or a weak fault may be masked).
  4. Remove the pack and examine the indicator sheet immediately.

4Interpreting the result

AppearanceInterpretationAction
Uniform colour change across the whole sheetPass — air removal and steam penetration are adequateRecord and release sterilizer for use
Lighter or unchanged area in the centreFail — residual air or non-condensable gasesTake the sterilizer out of service; call for maintenance; repeat test after repair
Uneven, patchy changeSuspect — possible wet steam or superheatingInvestigate steam quality; repeat test

5Documentation

File each test sheet, or a copy, with the date, sterilizer number, cycle number and operator. Most accreditation schemes require the Bowie-Dick record to be retained with the sterilizer log.

6Conclusion

The Bowie-Dick test takes ten minutes each morning and catches the single most common cause of steam sterilization failure — trapped air. Combined with a Type 1 tape on every pack, a Type 4/5 indicator inside every pack and routine biological monitoring, it forms the backbone of a compliant monitoring programme.

7Frequently asked questions

Does a gravity-displacement sterilizer need a Bowie-Dick test?

No. The test is specific to sterilizers with active air removal (pre-vacuum). Gravity sterilizers are monitored with chemical and biological indicators only.

Can I run the Bowie-Dick test with a load in the chamber?

No. The chamber must be empty apart from the test pack, so that the test reflects the sterilizer's own performance.