1What does VHP mean?

VHP stands for vaporized hydrogen peroxide. Liquid hydrogen peroxide at around 3 % is only a disinfectant, but when concentrated peroxide is vaporized under vacuum it becomes a powerful sterilant that diffuses over surfaces and into lumens, oxidizing cell membranes, enzymes and DNA.

2Hydrogen peroxide gas plasma vs VHP

Hydrogen peroxide gas plasma is created during the plasma phase, after peroxide diffusion, when radio-frequency energy excites the vapour into a cloud of ions, free radicals and UV photons. Most of the lethality comes from the peroxide vapour itself; the plasma phase mainly dissociates residues so the load comes out dry and safe. Non-plasma VH2O2 sterilizers achieve the same result with vacuum and catalytic destruction. The international standard for these processes is ISO 22441.

3Typical cycle

  1. Vacuum to remove air and moisture.
  2. Injection and vaporization of H₂O₂.
  3. Diffusion through packaging into the load.
  4. Plasma or vent phase to remove residue.
  5. Repeat (most cycles have two half-cycles).

Cycle times range from about 28 minutes for standard loads to 75 minutes for long-lumen or flexible-endoscope cycles, at 45–55 °C.

4What can and cannot be processed

CompatibleNot compatible
Rigid and semi-rigid endoscopes, cameras, light cablesPaper, cotton, linen, gauze (cellulose absorbs H₂O₂)
Electrosurgical and battery-powered devices (per IFU)Liquids, powders, oils
Stainless steel and most plasticsLong, narrow lumens beyond the validated limits
Ophthalmic and microsurgical instrumentsItems that cannot be dried completely

Loads must be thoroughly dry — residual water causes cycle cancellations.

5Packaging for plasma sterilization

Use Tyvek®/film pouches and reels or polypropylene (SMS) wrap and non-cellulose trays. Standard medical paper pouches cannot be used. Kangde Med's plasma indicator tape is designed for the Tyvek® and polypropylene materials used in these cycles. Read Tyvek pouches vs paper pouches.

6Monitoring VH2O2 cycles

  • External — H₂O₂ process indicator tape: Kangde Med plasma tape changes from blue to pink/red.
  • Internal — a Type 4 plasma indicator strip, heavy-metal-free, placed inside each pouch or tray.
  • Biological — self-contained BIs with *G. stearothermophilus* spores, daily or per manufacturer guidance.

7VHP decontamination vs sterilization

VHP is also used to decontaminate rooms, isolators and biosafety cabinets ("VHP decon"). Those processes achieve high-log bioburden reduction on exposed surfaces but are not a substitute for packaged-device sterilization.

8Choosing a low-temperature method

Compare VH2O2 with ethylene oxide: EO penetrates better and tolerates paper but needs long aeration; VH2O2 is fast and residue-free but has material and lumen limits. Many hospitals run both.

9Frequently asked questions

What temperature is VHP sterilization?

Most hydrogen peroxide sterilizers operate at about 45–55 °C, which is why they are classed as low-temperature sterilization.

Can paper sterilization pouches be used in plasma sterilizers?

No. Cellulose absorbs hydrogen peroxide and causes cycle aborts. Use Tyvek®/film pouches or polypropylene wrap.

How long does hydrogen peroxide plasma sterilization take?

About 28 to 75 minutes, depending on the sterilizer and cycle selected for the load.

Is VHP safe?

VH2O2 is toxic during the cycle but breaks down into water and oxygen; there are no toxic residues to aerate, unlike ethylene oxide.

What colour does a plasma indicator change?

It depends on the ink. Kangde Med plasma tape changes from blue to pink/red; always compare with the reference colour supplied.