1Core test methods

TestStandardWhat it measuresTypical use
Seal strength (peel)ASTM F88 / EN 868-5 Annex DForce to separate the sealProcess development and lot release
Dye penetrationASTM F1929Channels in porous sealsIntegrity verification
Bubble emissionASTM F2096Gross leaks under pressureNon-porous and some porous packs
Burst / creepASTM F1140, F2054Overall package strengthProcess monitoring
Visual inspectionASTM F1886Seal defects ≥75 µm100 % or sampling
Microbial barrierISO 11607 Annex / ASTM F1608Material barrierMaterial qualification

2ASTM F88 in practice

Strips 25 mm wide are cut across the seal and pulled apart in a tensile tester at a set speed (often 200–300 mm/min). Three tail-holding techniques exist (unsupported, supported at 90°, supported at 180°); the report should state which. Acceptance criteria are defined by the manufacturer — ASTM F88 itself does not set a pass value — but EN 868-5 sets a minimum of 1.2 N/15 mm for paper/film pouches and reels, and many specifications add an upper limit so the pack still peels cleanly.

3Integrity testing

Strong seals can still have channels. ASTM F1929 applies a low-surface-tension dye to the seal edge; any wicking through the seal within 5 seconds indicates a channel. This is particularly useful for paper/film and Tyvek® pouches.

4Building a packaging validation plan (ISO 11607)

  1. Qualify materials: microbial barrier, sterilant compatibility, biocompatibility.
  2. Validate the sealing process (IQ/OQ/PQ) — see medical heat sealers.
  3. Sterilize worst-case packages (e.g. double cycles).
  4. Age them — accelerated aging per ASTM F1980 (e.g. 55 °C) plus real-time aging.
  5. Simulate distribution (ASTM D4169 / ISTA).
  6. Test strength and integrity; document.

5What to ask your packaging supplier

Seal-strength data per lot, dye-penetration results, sterilization compatibility statements, aging data, and certificates of conformity. Kangde Med provides EN 868-5 and ISO 11607-1 documentation for pouches and reels and roll-stock paper and films.

6Frequently asked questions

What is ASTM F88?

ASTM F88 is the standard test method for seal strength of flexible barrier materials, measuring the force needed to peel apart a 25 mm wide strip of seal.

What is the acceptance criterion for seal strength?

ASTM F88 does not set one; the manufacturer defines it. EN 868-5 requires at least 1.2 N/15 mm for paper/film pouches and reels.

What is the difference between seal strength and seal integrity?

Seal strength measures how much force the seal withstands; integrity confirms there are no channels or leaks that could let microorganisms in.

When should package testing be performed?

During validation after sterilization, aging and distribution simulation, and routinely for lot release and process monitoring.