1What is a biological indicator?
A biological indicator is a standardised challenge of living spores chosen because they are more resistant to the sterilization process than the bioburden normally found on instruments. If the cycle kills them, it can be assumed to have killed everything else. BIs are specified in the ISO 11138 series.
2Formats
| Format | How it works | Read time |
|---|---|---|
| Spore strips | Paper strip in a glassine envelope; transferred to growth medium after the cycle | 2–7 days (often via mail-in service) |
| Self-contained BI (SCBI) | Spores and medium in one vial; crushed after the cycle | 24–48 h; rapid readers 20 min–3 h |
| Process challenge device (PCD) | BI inside a test pack that simulates a hard-to-sterilize load | As per BI |
3Organisms by process
- Steam — *Geobacillus stearothermophilus* (incubated at 55–60 °C).
- Hydrogen peroxide — *G. stearothermophilus*.
- Ethylene oxide — *Bacillus atrophaeus* (37 °C).
- Dry heat — *B. atrophaeus*.
4How often should you spore test?
| Setting | Recommended frequency |
|---|---|
| Hospital steam sterilizers | At least weekly, preferably daily; every load with implants |
| Dental offices (CDC) | At least weekly; every implant load |
| After installation, relocation or major repair | Three consecutive empty-chamber tests (pre-vacuum) |
| EO sterilizers | Every load |
| VH2O2 | Daily or per manufacturer |
5How to run an autoclave spore test
- Place the BI (or PCD) in the most challenging spot of a typical load.
- Run the cycle normally.
- Activate and incubate the processed BI alongside an unprocessed control from the same lot.
- Control must grow; processed BI must not.
- Record results with sterilizer, date, load number and operator.
6What if a spore test fails?
Remove the sterilizer from service, recall items processed since the last negative BI as far as possible, review cycle records, chemical indicator results, loading and packaging, retest, and repair if needed. Release the sterilizer only after three negative tests (or as local guidance requires).
7Using CIs alongside BIs
Because a BI takes time to read, every pack still needs an internal chemical indicator. A Class 5 integrating indicator in the PCD lets many facilities release non-implant loads while the BI incubates. Read more in chemical vs biological indicators.
8Frequently asked questions
What is Geobacillus stearothermophilus used for?
It is the biological indicator organism for steam and hydrogen peroxide sterilization because its spores are highly resistant to moist heat and peroxide.
How often should a dental autoclave be spore tested?
The CDC recommends at least weekly and with every load that contains an implantable device.
What does a positive spore test mean?
Spores survived the cycle, which suggests a sterilization failure. The sterilizer must be taken out of service and the cause investigated before reuse.
Can chemical indicators replace spore tests?
No. Chemical indicators show that process conditions were met; only biological indicators directly demonstrate microbial kill.