1Why EO needs its own indicators

Steam indicators respond to moist heat; EO indicators respond chemically to ethylene oxide gas in the presence of humidity. Using a steam indicator in an EO load gives no useful information, so facilities that run both processes need clearly different colours to avoid mix-ups.

2EO indicator tape vs EO indicator strip

EO indicator tapeEO indicator strip
ISO 11140-1 typeType 1 process indicatorType 4 multi-variable indicator
PlacementOutside every pack — closes wraps and identifies processed packsInside the pack, in the hardest-to-reach area
What it showsThe pack went through an EO processEO concentration, temperature, RH and time reached stated values
Kangde Med colour changeOrange → greenRed → blue
ProductEO Indicator TapeEO Indicator Strip

Sterilization pouches and reels also carry a printed EO process indicator (pink → yellow on Kangde Med pouches) outside the seal area.

3How to read an EO indicator

  1. Remove the pack after aeration.
  2. Compare the indicator with the reference colour printed on the tape core, strip or box.
  3. A complete, uniform change = pass. A partial or uneven change = fail: quarantine the pack, review the cycle record and reprocess.
  4. Record the result with the load number.

4Common mistakes

  • Reading the strip before the cycle and aeration are complete.
  • Placing the strip in an easy spot rather than the centre of the densest area.
  • Storing indicators near EO cylinders, solvents or in direct sunlight, which can cause premature change.
  • Using expired product — always check the lot number and expiry date.

5Buying EO indicators

Ask for ISO 11140-1 test reports for each type, an ISO 13485 certificate, shelf-life data and samples to trial in your own sterilizer. Distributors should also confirm private-label options and carton-level lot traceability. Learn more about indicator classes in Class 1 to Class 6 explained and about the process itself in our ethylene oxide sterilization guide.

6Frequently asked questions

What is an EO indicator?

A chemical indicator whose ink changes colour when exposed to ethylene oxide gas, used to show that a pack was processed (tape) or that EO penetrated the pack (strip).

Does a changed EO indicator mean the pack is sterile?

No. It confirms exposure to the stated conditions. Sterility is supported by the combination of physical cycle records, chemical indicators and biological indicators.

Where should an EO indicator strip be placed?

Inside the pack, in the area most difficult for gas to reach — typically the centre of the densest part of the load or inside lumened device trays.

Can I use EO indicator tape in a steam autoclave?

No. Use steam indicator tape for steam cycles. EO inks are formulated for ethylene oxide and will not give a valid result in steam.